The ADI 1025 Bio Console

The basic Bio Console is configured to accept 3 fixed speed pumps. The pump heads that are added to these drives are peristaltic tubing pumps and are available in various sizes. The Bio Console accepts up to 4 rotameter flow tubes. The flow range and gas type can be specified for the application, as required. Each rotameter provides a needle valve for manual adjustment of gas flow. A separate 3-way valve allows the operator to select gas flow to the sparger or gas overlay.

A third position allows the gas flow to be shut off. In addition, solenoid valves are available for each rotameter to control gas flow as a function of dissolved oxygen or pH setpoint.

As an option the rotameter assemblies can be replaced with mass flow controllers. The Bio Console can accept up to 2 mass flow controllers, of various capacities. The ADI 1010 Bio Controller can control up to 4 mass flow controllers. Additional mass flow controllers can be mounted in a gas flow control cabinet.

The Bio Console also can incorporate a temperature controller. For temperature control, several options are available. A thermocirculator can be installed which allows water to flow to the heat exchanger or jacket of the bioreactor. As an alternative, a heating blanket or heating element can be plugged into the Bio Console. For cooling purposes, a cold water solenoid valve can be installed.

All devices described above are controlled from the ADI 1010 Bio Controller which is installed on top of the Bio Console.